Skip to content

Turn any AI agent into an AI Scientist. The #1 Agent Skills library for science, used by 250,000+ scientists worldwide. 177 ready-to-use validated skills plus 100+ scientific databases covering biology, chemistry, medicine, and drug discovery. Compatible with Cursor, Claude Code, Codex, Pi, Antigravity, and the open Agent Skills standard.

4.1 KB
Raw Blame

Analytical Procedure Validation Protocol

Every bracketed field is a decision to make and record before data collection. plan_validation.py --protocol generates a framework-specific version of this document with the required characteristics already listed.

Field Value
Protocol number / version [ ]
Analytical procedure identifier and version [ ]
Product / analyte / matrix [ ]
Measured quality attribute [ ] assay / impurity (quantitative) / impurity (limit) / identity / other
Governing framework and section [ ]
Regional expectation confirmed with [ ]
Related development report (ICH Q14) [ ]
Author / date [ ]
Technical reviewer / date [ ]
Quality unit approval / date [ ]

1. Intended purpose and analytical target profile

  • Measurand and reporting unit: [ ]
  • Decision the result supports: [ ] release / stability / in-process / clinical / other
  • Specification or reporting limits served: [ ]
  • Required reportable range, derived from the specification: [ ]
  • Performance characteristics and criteria (formal ATP optional under Q14): [ ]

2. Pre-stated acceptance criteria

State an appropriate criterion and its justification for every characteristic to be validated; identity or specificity criteria need not always be numeric. A criterion with no justification traceable to the specification, the ATP, or development data is not defensible.

Characteristic Criterion Justification Framework reference
Specificity / selectivity [ ] [ ] [ ]
Response (calibration model) [ ] [ ] [ ]
Lower range limit (DL / QL) [ ] [ ] [ ]
Accuracy [ ] [ ] [ ]
Repeatability [ ] [ ] [ ]
Intermediate precision [ ] [ ] [ ]
Combined accuracy and precision, if used [ ] [ ] [ ]
  • Interval to be reported alongside accuracy and precision: [ ] confidence level [ ]
  • Does the criterion apply to the point estimate or to the whole interval? [ ]

3. Study design

Characteristic Levels Replicates Runs / days / analysts / instruments
Response [ ] (minimum 5 for ICH Q2(R2)) [ ] [ ]
Accuracy [ ] [ ] [ ]
Repeatability [ ] [ ] [ ]
Intermediate precision [ ] [ ] [ ]
Lower range limit [ ] [ ] [ ]
  • Replicate count matches the routine reportable result: [ ] yes / [ ] justified deviation: [ ]
  • Calibration model and weighting, fixed in advance: [ ] unweighted / 1/x / 1/x² / non-linear / multivariate
  • Randomisation and run order: [ ]
  • Independent preparation, sample, group, and injection identifiers: [ ]
  • CI unit of independence and handling of unbalanced groups: [ ]
  • Prior knowledge or development data used in place of a test, with justification: [ ]

4. Materials

Item Identity / grade Lot Assigned value and uncertainty Expiry
Reference material [ ] [ ] [ ] [ ]
Impurity standards [ ] [ ] [ ] [ ]
Blank / placebo matrix [ ] [ ] — [ ]

5. Sample and solution handling

  • Preparation procedure and dilution scheme: [ ]
  • Solution stability window to be demonstrated: [ ]
  • Storage conditions: [ ]

6. Specificity and stability-indicating properties

  • Interferences to be challenged: [ ]
  • Forced degradation conditions, if a stability-indicating claim is made: [ ]
  • Orthogonal procedure, if used, and its accuracy: [ ]

7. Robustness (normally development, ICH Q14)

Parameter Nominal Range varied Effect assessed on
[ ] [ ] [ ] [ ]

8. Statistical treatment

  • Software, version, and how calculations are verified: [ ]
  • Handling of outliers, stated in advance: [ ]
  • Scripts to be used and their output retained as records: [ ]

9. Deviations and data integrity

  • Deviation identification, assessment and approval route: [ ]
  • All results will be reported, including out-of-criteria values: [ ] confirmed
  • Raw data location, audit trail, and review: [ ]

10. Approvals

Role Name Signature Date
Author
Technical reviewer
Quality unit