Turn any AI agent into an AI Scientist. The #1 Agent Skills library for science, used by 250,000+ scientists worldwide. 177 ready-to-use validated skills plus 100+ scientific databases covering biology, chemistry, medicine, and drug discovery. Compatible with Cursor, Claude Code, Codex, Pi, Antigravity, and the open Agent Skills standard.
Python agent-skillsai-scientistbioinformaticschemoinformaticsclaudeclaude-skillsclaudecodeclinical-researchcomputational-biologydata-analysisdrug-discoverygenomicsmaterials-sciencemetabolomicsproteomicsscientific-computingscientific-visualization
#!/usr/bin/env python3
"""Apply ICH M10 acceptance criteria to a bioanalytical run or an ISR dataset.
Chromatographic assays and ligand binding assays carry DIFFERENT numeric
criteria in ICH M10, and conflating them is the most common error in this area.
--modality is therefore mandatory: nothing here has a default.
# calibration standards and QCs from one analytical run
python3 check_bioanalytical_run.py --modality chromatographic --run run1.csv
# incurred sample reanalysis
python3 check_bioanalytical_run.py --modality lba --isr isr.csv
# LBA total-error criterion from accuracy/precision validation data
python3 check_bioanalytical_run.py --modality lba --total-error ap.csv
Input for --run: columns `type` (calibrator|qc), `nominal`, `measured`, and an
optional `label` (e.g. LLOQ, low, medium, high, ULOQ, or ANCHOR for LBA anchor
points, which are excluded from the calibration pass count).
Input for --isr: columns `original` and `repeat`.
Input for --total-error: columns `label`, `accuracy_pct` (signed percent bias,
not recovery), `precision_pct` (non-negative CV).
Exit codes: 0 all applied criteria met, 1 criteria not met, 2 bad input.
"""
from __future__ import annotations
import argparse
import math
import sys
sys.path.insert(0, str(__import__("pathlib").Path(__file__).resolve().parent))
from _catalog import M10_CRITERIA # noqa: E402
from _common import ( # noqa: E402
EXIT_FINDINGS,
EXIT_OK,
InputError,
add_common_args,
emit,
finding,
note,
parse_rows,
read_input,
require_columns,
run_cli,
to_float,
)
LIMIT_LABELS = {"lloq", "uloq"}
def calibrator_tolerance(label: str, crit: dict) -> float:
"""Tolerance for a calibration standard at this label.
Routine-run QCs are not routed through here: M10 applies one flat tolerance
to every QC level for run acceptance, which check_run uses directly.
"""
low = label.strip().lower()
if low == "lloq":
return crit["calibration_tolerance_lloq_pct"]
if low == "uloq":
return crit["calibration_tolerance_uloq_pct"]
return crit["calibration_tolerance_pct"]
def check_run(path: str, crit: dict) -> tuple[list[dict], list[str]]:
rows = parse_rows(read_input(path), path)
require_columns(rows, ["type", "nominal", "measured"])
findings: list[str] = []
detail: list[dict] = []
cal_pass = cal_total = 0
cal_by_level: dict[float, list[bool]] = {}
qc_by_level: dict[float, list[bool]] = {}
qc_labels: dict[float, str] = {}
qc_pass = qc_total = 0
for i, r in enumerate(rows):
kind = (r["type"] or "").strip().lower()
label = (r.get("label") or "").strip()
nominal = to_float(r["nominal"], "nominal", i)
measured = to_float(r["measured"], "measured", i)
if nominal <= 0:
raise InputError(f"row {i + 1}: nominal must be positive")
dev = 100.0 * (measured - nominal) / nominal
if kind in ("calibrator", "cal", "standard", "std"):
if label.lower() == "anchor":
if crit["total_error_pct"] is None:
raise InputError("ANCHOR exclusion is supported for LBA only")
detail.append({"type": "calibrator", "label": "ANCHOR", "nominal": nominal,
"measured": measured, "deviation_pct": dev,
"tolerance_pct": float("nan"), "within": "excluded"})
continue
tol = calibrator_tolerance(label, crit)
ok = abs(dev) <= tol
cal_total += 1
cal_pass += int(ok)
cal_by_level.setdefault(nominal, []).append(ok)
detail.append({"type": "calibrator", "label": label or "-", "nominal": nominal,
"measured": measured, "deviation_pct": dev,
"tolerance_pct": tol, "within": ok})
elif kind in ("qc", "quality-control"):
tol = crit["qc_run_tolerance_pct"]
ok = abs(dev) <= tol
key = nominal
qc_labels[key] = label or f"{nominal:g}"
qc_by_level.setdefault(key, []).append(ok)
qc_total += 1
qc_pass += int(ok)
detail.append({"type": "qc", "label": qc_labels[key], "nominal": nominal,
"measured": measured, "deviation_pct": dev,
"tolerance_pct": tol, "within": ok})
else:
raise InputError(
f"row {i + 1}: type must be calibrator or qc, got {r['type']!r}"
)
# Calibration curve criteria.
if cal_total:
n_levels = sum(any(flags) for flags in cal_by_level.values())
need_levels = crit["calibration_min_levels"]
if n_levels < need_levels:
findings.append(
f"calibration curve has {n_levels} passing concentration levels; M10 requires a "
f"minimum of {need_levels} passing levels"
)
frac = cal_pass / cal_total
need = crit["calibration_min_pass_fraction"]
if frac < need:
findings.append(
f"{cal_pass}/{cal_total} calibration standards within tolerance "
f"({frac * 100:.1f}%); M10 requires at least {need * 100:.0f}%"
)
for nominal, flags in cal_by_level.items():
if sum(flags) / len(flags) < 0.5:
findings.append(f"calibration level {nominal:g}: fewer than 50% of replicates pass")
if cal_pass < cal_total:
findings.append("failed calibrators require documented exclusion and a refitted curve; "
"this checker does not refit or establish a revised range")
for boundary in ("lloq", "uloq"):
points = [r for r in detail if r["type"] == "calibrator" and r["label"].lower() == boundary]
if not points:
findings.append(f"label {boundary.upper()} explicitly to verify boundary tolerances")
else:
expected = min(cal_by_level) if boundary == "lloq" else max(cal_by_level)
if any(point["nominal"] != expected for point in points):
raise InputError(f"{boundary.upper()} label does not identify the calibration boundary")
for anchor in (r for r in detail if r["within"] == "excluded"):
if min(cal_by_level) <= anchor["nominal"] <= max(cal_by_level):
raise InputError("LBA anchor points must lie outside the quantitation range")
else:
findings.append("no non-anchor calibration standards supplied; run assessment incomplete")
# Routine run QC criteria: both the overall fraction and per-level fraction.
if qc_total:
n_qc_levels = len(qc_by_level)
if n_qc_levels < crit["qc_levels_routine_run"]:
findings.append(
f"{n_qc_levels} QC levels present; M10 expects at least "
f"{crit['qc_levels_routine_run']} for run acceptance"
)
frac = qc_pass / qc_total
need = crit["qc_run_pass_fraction"]
if frac < need:
findings.append(
f"{qc_pass}/{qc_total} QCs within +/-{crit['qc_run_tolerance_pct']:.0f}% "
f"({frac * 100:.1f}%); M10 requires at least {need * 100:.0f}% of the total"
)
for level, flags in sorted(qc_by_level.items()):
if len(flags) < 2:
findings.append(f"QC level {qc_labels[level]}: at least two QC samples are needed")
level_frac = sum(flags) / len(flags)
if level_frac < crit["qc_run_pass_fraction_per_level"]:
findings.append(
f"QC level {qc_labels[level]}: {sum(flags)}/{len(flags)} within tolerance "
f"({level_frac * 100:.0f}%); M10 requires at least "
f"{crit['qc_run_pass_fraction_per_level'] * 100:.0f}% at each level"
)
else:
findings.append("no QCs supplied; run assessment incomplete")
return detail, findings
def check_isr(path: str, crit: dict) -> tuple[list[dict], list[str]]:
rows = parse_rows(read_input(path), path)
require_columns(rows, ["original", "repeat"])
tol = crit["isr_tolerance_pct"]
detail, findings = [], []
passes = 0
for i, r in enumerate(rows):
original = to_float(r["original"], "original", i)
repeat = to_float(r["repeat"], "repeat", i)
if original <= 0 or repeat <= 0:
raise InputError("ISR requires positive quantified concentrations within the validated range")
mean_val = 0.5 * (original + repeat)
if mean_val == 0:
raise InputError(f"row {i + 1}: mean of original and repeat is zero")
# M10 defines the ISR percent difference against the mean of the two.
diff = 100.0 * (repeat - original) / mean_val
ok = abs(diff) <= tol
passes += int(ok)
detail.append({"sample": r.get("sample", str(i + 1)), "original": original,
"repeat": repeat, "mean": mean_val, "percent_difference": diff,
"tolerance_pct": tol, "within": ok})
frac = passes / len(rows)
if frac < crit["isr_pass_fraction"]:
findings.append(
f"ISR: {passes}/{len(rows)} repeats within +/-{tol:.0f}% ({frac * 100:.1f}%); "
f"M10 requires at least {crit['isr_pass_fraction'] * 100:.0f}%"
)
return detail, findings
def check_total_error(path: str, crit: dict) -> tuple[list[dict], list[str]]:
if crit["total_error_pct"] is None:
raise InputError(
"ICH M10 states a total-error criterion for ligand binding assays only; "
"use --modality lba, or omit --total-error for a chromatographic assay"
)
rows = parse_rows(read_input(path), path)
require_columns(rows, ["label", "accuracy_pct", "precision_pct"])
detail, findings = [], []
for i, r in enumerate(rows):
label = (r["label"] or "").strip()
acc = to_float(r["accuracy_pct"], "accuracy_pct", i)
prec = to_float(r["precision_pct"], "precision_pct", i)
if prec < 0:
raise InputError("precision_pct must be a non-negative CV; accuracy_pct is signed bias, not recovery")
total = abs(acc) + abs(prec)
limit = (
crit["total_error_pct_at_limits"]
if label.lower() in LIMIT_LABELS
else crit["total_error_pct"]
)
ok = total <= limit
detail.append({"label": label or "-", "accuracy_pct": acc, "precision_pct": prec,
"total_error_pct": total, "limit_pct": limit, "within": ok})
if not ok:
findings.append(
f"{label or 'level ' + str(i + 1)}: total error {total:.2f}% exceeds "
f"{limit:.0f}%"
)
return detail, findings
def main() -> int:
parser = argparse.ArgumentParser(
description="Apply ICH M10 acceptance criteria to bioanalytical data."
)
parser.add_argument("--modality", required=True, choices=sorted(M10_CRITERIA),
help="chromatographic or lba -- the criteria differ and there is no default")
parser.add_argument("--run", help="calibrators and QCs from an analytical run")
parser.add_argument("--isr", help="incurred sample reanalysis data")
parser.add_argument("--total-error", dest="total_error",
help="accuracy/precision per level (ligand binding assays only)")
parser.add_argument("--criteria", action="store_true",
help="print the criteria that would be applied and exit")
add_common_args(parser)
args = parser.parse_args()
crit = M10_CRITERIA[args.modality]
if args.criteria:
rows = [
{"criterion": k, "value": ("n/a" if v is None else v)}
for k, v in crit.items()
if k not in ("label", "notes")
]
emit(rows, args.format)
note(crit["label"])
note(crit["notes"])
return EXIT_OK
if not any((args.run, args.isr, args.total_error)):
raise InputError("supply at least one of --run, --isr, --total-error, or --criteria")
detail: list[dict] = []
findings: list[str] = []
sections: dict[str, list[dict]] = {}
if args.run:
d, f = check_run(args.run, crit)
sections["run"] = d
findings += f
if args.isr:
d, f = check_isr(args.isr, crit)
sections["isr"] = d
findings += f
if args.total_error:
d, f = check_total_error(args.total_error, crit)
sections["total_error"] = d
findings += f
if args.format == "json":
emit([{"modality": args.modality, **sections, "findings": findings}], "json")
else:
for name, rows in sections.items():
print(f"[{name}]")
emit(rows, args.format)
print()
note(crit["label"])
note(crit["notes"])
note(
"chromatographic and ligand binding assay criteria differ throughout M10; this run was "
f"assessed as: {args.modality}"
)
for f in findings:
finding(f)
if not findings:
note("all applied criteria met")
note(
"run acceptance is a documented decision by the analyst; this tool applies stated "
"criteria and does not accept or reject a run"
)
if args.run:
note("one row must represent a reportable standard or QC sample, not an individual LBA well; "
"check blanks, zero samples, study-sample-dependent 5% QC count, bracketing, "
"plate/batch criteria, and calibration refitting separately")
return EXIT_FINDINGS if findings else EXIT_OK
if __name__ == "__main__":
run_cli(main)