Turn any AI agent into an AI Scientist. The #1 Agent Skills library for science, used by 250,000+ scientists worldwide. 177 ready-to-use validated skills plus 100+ scientific databases covering biology, chemistry, medicine, and drug discovery. Compatible with Cursor, Claude Code, Codex, Pi, Antigravity, and the open Agent Skills standard.
Signal processing (pyOpenMS 3.6.0)
Decide what the signal represents
Use spec.getType() and acquisition/conversion provenance. Its enum is
ms.SpectrumSettings.SpectrumType.PROFILE, CENTROID, or UNKNOWN.
isSorted() tests m/z ordering only. getType(True) can estimate type from the
signal when metadata is insufficient, but an estimate is not acquisition proof.
process_spectra.py applies smoothing, picking, normalization, S/N filtering,
then intensity thresholding to selected MS levels; chromatograms pass through.
It rejects centroid spectra for smoothing/picking and requires
--assume-profile for independently verified unknown-type spectra.
The native tests check a synthetic Gaussian profile, selected-level isolation,
metadata/peak-array retention and writeback of filtered spectra.
python scripts/process_spectra.py profile.mzML picked.mzML --ms-level 1 --pick
python scripts/process_spectra.py spectra.mzML scaled.mzML --ms-level 2 --normalize to_one --threshold 0.01
These are file-dependent templates. Parameters need representative instrument
QC; smoothing can distort narrow peaks. PeakPickerHiRes is for well sampled
high-resolution profile peaks. Do not repeatedly centroid already picked data.
API patterns
import pyopenms as ms
import numpy as np
spec = ms.MSSpectrum()
spec.setType(ms.SpectrumSettings.SpectrumType.PROFILE)
mz = np.linspace(499.9, 500.1, 101)
spec.set_peaks((mz, 1000 * np.exp(-0.5 * ((mz - 500) / 0.01) ** 2)))
spec.sortByPosition()
picker = ms.PeakPickerHiRes()
picked = ms.MSSpectrum()
picker.pick(spec, picked)
assert len(picked) == 1
assert abs(picked[0].getMZ() - 500) < 0.001
GaussFilter.filter(spec) uses gaussian_width in m/z units for a
spectrum; use_ppm_tolerance selects ppm_tolerance instead. Chromatogram
widths use seconds. The 3.6 filterSpectrum compatibility alias is broken for these two smoothers;
use their native filter(spec) method. SavitzkyGolayFilter uses a point-count frame_length and
polynomial_order; validate sampling regularity and window size.
PeakPickerIterative has signal_to_noise_, peak_width, and
nr_iterations_ parameters; this alternative was signature/parameter checked,
not benchmarked on low-resolution instrument data.
Normalization (Normalizer.filterSpectrum) supports to_one and to_TIC.
Thresholds afterwards are on normalized intensities; a threshold of 10 after
unit-TIC normalization removes all nonnegative peaks. Within-scan normalization
usually destroys the abundance information needed for label-free quantification.
For filtering, use spec.select(indices) to keep float/integer/string data
arrays synchronized. set_peaks() alone does not filter associated arrays.
Spectra read with getSpectrum, indexing or iteration may be copies: explicitly
replace the full spectrum list with exp.setSpectra(processed).
Other algorithms and alignment
ThresholdMower, WindowMower, NLargest, and MorphologicalFilter expose
parameters through getParameters(); inspect their installed filterSpectrum
signatures before use. Their particular noise/baseline assumptions are not
validated by a successful call. For SpectraMerger, block-wise merging uses
block_method:* options; setting average_gaussian:* while calling
mergeSpectraBlockWise does not configure Gaussian averaging.
Deisotoper.deisotopeAndSingleCharge acts on a sorted centroided spectrum;
record tolerance units, charge range, isotope count, and whether intensity is
summed or masses are converted to singly charged ions. This is not calibration.
For feature deconvolution use a FeatureMap and FeatureDeconvolution.compute
with output map/groups/edges; there is no potential_charge_states parameter.
For RT alignment, call aligner.setReference(reference_map),
aligner.align(sample_map, trafo), then
MapAlignmentTransformer().transformRetentionTimes(sample_map, trafo, True).
See the feature reference and bundled alignment script. Retention alignment
does not correct mass errors.
Calibration: verified signature, illustrative workflow
InternalCalibration.calibrate(exp, reference_masses) is not a valid API.
First collect calibrants using fillCalibrants, then fit/apply a model. The
following is a template requiring independently established ion m/z values and
a representative run; it was not validated on instrument data:
calibration = ms.InternalCalibration()
locks = [ms.InternalCalibration_LockMass(500.0, 1, 1),
ms.InternalCalibration_LockMass(1000.0, 1, 1)]
found, failed = calibration.fillCalibrants(exp, locks, 20.0, False, False, False)
if found < 2:
raise ValueError("Insufficient calibrants; do not fit")
ok = calibration.calibrate(exp, [1], ms.MZTrafoModel.MODELTYPE.LINEAR,
-1.0, False, 5.0, 5.0)
if not ok:
raise ValueError("Calibration residual criteria failed")
Lock masses include adducts and require correct MS level and charge. Two points
are merely a minimal algebraic requirement for a line; use adequate mass/RT
coverage, independent validation ions, outlier diagnostics, isotope checks and
before/after ppm residuals. Do not calibrate on unverified accurate-mass database
candidates and then use the reduced error as independent identification evidence.
Negative rt_chunk selects one global model; positive windows are seconds.
R is needed only for requested calibration plot output, not this no-plot template.
Sources: InternalCalibration release source, OpenMS changes, and installed 3.6.0 binding docstrings/defaults.